Nitrogen is a lead parameter to determine surface water quality and wastewater effluent values. Both, organic and inorganic substances contribute to the total nitrogen load. Increased nitrogen contents indicate influences from wastewater, landfill leachate or agriculture. Liquiline System CA80TN analyzer offers precise online measurement of total nitrogen and ensures highest process safety. Advanced diagnostics with remote access support you in providing process documentation to the authorities.
The Liquiline System TN analyzer monitors organic (e.g. proteins, urea) and inorganic (e.g. nitrate, nitrite, ammonium) nitrogen compounds in the following applications:
Inlet and outlet of wastewater treatment plants for documentation and calculation of the cleaning capacity
Discharges of industrial wastewater treatment plants to determine discharge fees and to support the polluter-pays-principle
Monitoring of surface water quality
Benefícios
Standardized alkaline persulfate digestion and UV-photometric measurement according to HJ636 offer direct comparability to most cuvette tests.
Robust: Titanium reactor with sapphire glass windows ensures a long lifetime.
Flexible: Integrated dilution module covers a wide measuring range.
Fast and easy process integration: Direct installation of self-priming version or y-strainer for bypass applications.
Easy upgrade to a complete measuring station - simply by adding modules and connecting Memosens sensors.
Drive connectivity: Liquiline Edge Module CYY7 transfers process data to Netilion, independently of the fieldbus. This allows remote overview and the use of IIoT services. It’s available in new products or as upgrade to installed Liquiline platforms.
Vídeos
Memosens 2.0: Tecnologia inovadora de sensor aprimorada
Facilite a vida, aumente a segurança de processos e permaneça conectado com a tecnologia digital Memosens 2.0.
Liquiline: plataforma de ponta do transmissor
Ao escolher a plataforma Liquiline, você opta por gastar menos dinheiro na manutenção do estoque, economizar tempo de instalação e obter segurança da operação!
A solução de monitoramento proporciona à piscicultura de Visp um controle estável e confiável do oxigênio, garantindo condições ideais para os peixes e reduzindo o trabalho e os riscos.
Proline Teqwave MW 300 delivers stable real-time data taken directly from the process, reduces maintenance work, and allows quicker intervention if changes to the process are required.
Kerry DCF produces spreads and butter products for the retail market. The company wanted to automate its test for chemical oxygen demand in its treated effluent to save time and optimise the process.
O Memosens CPS11E garante um monitoramento preciso do pH na produção de biodiesel, melhorando a qualidade do produto, a estabilidade do processo e a vida útil do sensor em condições adversas.
Process analyzer for total nitrogen in aqueous solutions UV photometric method following HJ636-2012
Size
Housing (open version): 793 x 530 x 417 mm 31.22 x 20.87 x 16.42 in Housing (closed version): 793 x 530 x 463 mm 31.22 x 20.87 x 18.23 in Housing with base: 1723 x 530 x 463 mm 67.83 x 20.87 x 18.23 in
Design
Open design, cabinet and stand housing High-Performance plastic ASA-PC, additional stand coated steel
Process temperature
4 to 40 °C (39.2 to 104 °F)
Ambient temperature
5 to 40 °C (41 to 104 °F)
Process pressure
At atmospheric pressure, < 0.2 bar
Sample flow rate
Min. 15 ml/min (0.51 fl.oz/min)
Consistency of the sample
Aqueous, homogeneous
Application
Monitoring of wastewater treatment plant outlet Monitoring of the surface water quality Monitoring of industrial water treatment Monitoring and optimization of industrial wastewater treatment plants
Power supply
100 to 120 VAC / 200 to 240 VAC ± 10% 50 ± 1 or 60 ± 1,2 Hz
Output / communication
2x 0/4 to 20 mA Modbus RS485/TCP (optional) Webserver (optional) Ethernet/IP Profibus DP Alarm relay
Input
1 measuring channel 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measuring range
0 to 20 mg/l N 0 to 50 mg/l N 0 to 250 mg/l N
Consumables
Reagents and standard solutions CY80TN as well as cleaner CY800 are necessary for the operation Regular maintenance is done with the parts of the maintenance kit CAV880
Ensure optimal CA80 analyzer performance with expert commissioning and first year preventive maintenance service for seamless operation and reduced downtime